CODEL Operation and Maintenance Manual
Issue : A Rev. : 3 Date : 13/11/2018 Ref. : 100351
P a g e | 17
5.3.2 Detector levels
From the menu position below the detector levels can be observed. In zero conditions the E-Cell detector
levels should be:
Approx. 26215 for CO (cell 1)
Approx. 26215 for NO (cell 2)
Approx. 32767 for NO
2
(cell 3)
The values that are given relate to the default starting values for the E-Cell channels.
From the main display press or to reach Diagnostics. Press to access Diagnostic data.
Use the or key to reach E-Cell CH1, and press to access.
Use the or to reach DCal, and press to access.
Dcal Data figures should appear on the screen as shown below. This value shows the calibrated detector
values.
In order to access the detector level reading for E-Cell 2 & 3, follow the pathway above detailed in E-Cell
module comunication, E-Cell CH1 CO, but enter the relevant paths in the ‘select channel’ menu: