CODEL Operation and Maintenance Manual
Issue : A Rev. : 3 Date : 13/11/2018 Ref. : 100351
P a g e | 16
NOTE: If the display unit is unable to communication with the attached peripheral the screen will display a
communications fault. This will show an “E-Cell Comm Fault” or “Vis Comm Fault” respective of where the
fault resides.
If no devices are configured or have been disabled, the screen will display a warning message to alert the user
that no devices are attached.
5.3 E-Cell commissioning
5.3.1 E-Cell module communication
To ensure that there is a working communication between the display unit and the electrochemical cell
module, follow the steps below using the display.
From the main display press the or keys to reach Diagnostics. Press to access Diagnostic data.
Use the or keys to reach E-Cell CH1, and press to access.
Use the or keys to navigate to DCal, and press to access.
Data figures should appear on the screen as shown below. This indicates that the display module is
communicating with E-Cell module and is operating correctly.