Using the or keys, manipulates the onscreen value with the key moving to the next available
location or setting the entered value if all 5 fields are complete.
5.4.3 mA Span Value
When setting the Span value on the mA board, the menu system uses the measurand to determine how to
handle the requested information.
Example:
CO, NO, NO2, and the N/A measurands are all retrieved and set to 1 decimal place whereas VIS percent is set to
2 decimal places and VIS K to 4 decimal places.
To ensure the user is on the correct menu, the menu also displays the current measurand being worked on.
By pressing the key, the user can adjust the current value for the selected mA channel.
Using the or keys, manipulates the onscreen value with the key moving to the next available
location or setting the entered value if all 5 fields are complete.
NOTE: The value displayed on the screen is taken directly fom the mA EEPROM span location, when
changing meaurands it is important to remember to change the span factor.