TABLES 2 & 3: RECOMMENDATIONS AND DECLARATION
BY THE MANUFACTURER 
ELECTROMAGNETIC IMMUNITY
The Compex Edge is designed for use in the electromagnetic environment stipulated
below. The customer or the user of the Compex Edge must ensure that it is used in
this recommended environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
ELECTROMAGNETIC COMPATIBILITY
Immunity
Test
IEC 
Test Level
Compliance
Level
Electromagnetic
Environment 
Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV at the
contact
±16 kV in air
±8 kV at the
contact
±16 kV in air
Floors must be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at a minimum of 30%.
Electrical fast
transient/
burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not Applicable
System battery
powered
The quality of the power
supply should be that of
a typical commercial or
hospital environment.
Surge
(1)
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
Not Applicable
System battery
powered
The quality of the power
supply should be that of
a typical commercial or
hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec
Not Applicable
System battery-
powered
The quality of the power
supply should be that of
a typical commercial or
hospital environment. If
the Compex Sport Elite
user requires continuous
operation during mains
power cuts, we recommend
that the Compex Sport Elite
is powered by a UPS or a
battery.
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m Magnetic fields at the mains
frequency should be at the
level of a representative
site located in a typical
commercial or hospital
environment.