32 33
TESTING
C
ONTOUR
®
NEXT ONE blood glucose monitoring system
Control Solution
WARNING
• Shake the control solution well before testing.
CAUTION: Use only CONTOUR NEXT control
solution (Normal, Low, and High) with
your
C
ONTOUR NEXT ONE blood glucose
monitoring system. Using anything other than
C
ONTOUR NEXT control solution can cause
inaccurate results.
Quality Control
• Shake the control solution bottle well, about
15 times before every use.
• Unmixed control solution may cause inaccurate
results.
You should perform a control test when:
• Using your meter for the first time.
• You open a new bottle or package of test strips.
• You think your meter may not be working properly.
• You have repeated, unexpected blood sugar results.
WARNING
• Do not calibrate your continuous glucose monitoring device
from the control result.
• Do not calculate a bolus based on a control result.
CAUTION: Do not use expired materials.
Using expired material can cause inaccurate
results. Always check the expiration dates on
your test materials.
Normal, Low, or High control solutions are available and sold
separately if not included in the meter kit. You can test your
C
ONTOUR NEXT ONE meter with control solution when the temperature
is 15°C–35°C.
Always use C
ONTOUR NEXT control solutions. Other brands could yield
incorrect results. If you need help locating C
ONTOUR NEXT control
solutions, contact Diabetes Care Customer Service. See the back
cover for contact information.
Control Solution Testing
1. Remove a CONTOUR NEXT test strip from the bottle.
2. Tightly close the bottle lid immediately after you remove the
test strip.
3. Hold the test strip with the grey square end facing up.