Installation & User Guide SonoSelect and SonoSafe
2.8 Battery
1.
Before handling
module PCB ensure
relevant ESD regu-
lations are observed
(IEC 61340-5-1).
2.
Break installation
seal and open en-
closure.
3.
Disconnect battery
connector and
remove battery.
4.
Short circuit battery
connectors on PCB
using a small flat
head screwdriver.
5.
Connect new battery
to PCB.
6.
Fit battery
in enclosure.
7.
Close enclosure
making sure that no
cables interfere with
rubber sealing or
other internal parts.
8.
Confirm battery
change by pressing
the button 2 times
within 120 seconds.
Reset tamper alarm
(E13) via SonoApp
for SonoSelect 10.