Installation & User Guide SonoSelect and SonoSafe
1. Inside the box
Description of components included in the box
107/2015 © Danfoss VDIGH102
Installation Guide
SonoSelect and SonoSafe
energy meters
1 x
Energy meter
2 x
Gaskets
1 x
Temperature sensor
O-ring and mounting pin
4 x
Wired seals
1 x
Installation
& User Guide
Note!
For cooling meters a wall mounting kit is delivered with the product
2. Installation
2.1 Preparation
SonoSelect incorporates a tamper monitor-
ing function. If the calculator is opened the
meter will set alarm E13 in display.
Don’t open unless for adding communica-
tion module, replacing battery or installing
cables.
Reset requires Bluetooth dongle 014U1963
and SonoApp service tool.
SonoSelect is delivered with the option
of reconfiguring supply/return using the
Bluetooth dongle 014U1963 and SonoApp
service tool.
Note!
Product is approved for ambient temperature between 5-55 °C, non-condensing (indoor installation). It is
recommended to install Calculator at max. 45 °C to secure optimal conditions for battery lifetime. At media
temperatures below ambient temperature (cooling) the calculator must be mounted separately from the flow
sensor to prevent condensation.
Avoid installation stress from pipes and fittings.
Flush the system.