EasyManua.ls Logo

DayTronic 3000PLUS - Page 41

DayTronic 3000PLUS
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3KP64 - 3.17
3000PLUS WITH 5D64 INSTRUCTION MANUAL V. SB.2
SETUP STAGE 4(AV): ABSOLUTE VOLTAGE CALIBRATION
When the
3000PLUS
displays this . . . Do this . . .
ST
and
CL
are lit.
3. FRONT-PANEL CONFIGURATION
AND CALIBRATION
4(AV).a. Press
4(AV).b. The desired
FULL-SCALE OUTPUT IN ENGINEERING UNITS value
which you entered in Step 2.g will be displayed (having the decimal-
point precision specified in Step 2.d). If you now need to change this
number, you may use
and
as explained in Section 3.A to do so. Remember: Since
ABSOLUTE
VOLTAGE
calibration was selected in Step 4.b, this is the instrument
VOLTS reading that is to correspond to a full-scale analog output of
+5.000 volts.
Then press
NOTE: The 3000PLUS will now calculate a MODULE SCALING FAC-
TOR (MSF)
from the three numbers you have just entered (as
explained in Section 4.E and Appendix B, the
MSF is used as a gain
factor for the instruments full-scale input range). If this calculation
yields a gain that is too low (less than 1.0000, given the
FULL-SCALE
RANGE
value established in Step 2.b), you will see a display of
alternating with
If the calculated gain is too high (greater than 1.5999, given the current
RANGE value), you will see a display of
alternating with
Press
to acknowledge the high or low range condition.
If the 3000PLUS is able to reset the
RANGE value so that a legal
MSF may be calculated from the present calibration numbers, it will
display
alternating with
(contd)
ENTER
ENTER
ENTER
Daytronic Corporation
2211 Arbor Blvd. Dayton, OH 45439 (800) 668-4745
Tel: (937) 293-2566 Fax: (937) 293-2586 www.daytronic.com

Related product manuals