5. GETTING STARTED
Information #2
Instructions on properly mounting and preparing the detector system can be found in the
Technical Specications. Before operating the detector, read the complete documentation.
5.1. Startup Procedure
• Turn on the nitrogen or dry air ow at least 30 min before turning on the detector.
1
• Turn on the thermal stabilization unit and set the operation temperature as specied in the Technical
Specications. Please read the thermal stabilization unit manual, as some models must be powered and
additionally activated in order to operate properly.
2
• Connect all the required cables for power, data, and trigger signals (if applicable) to the detector and the
detector control unit.
• Turn on the detector.
• Turn on the detector control unit. Please allow 6 minutes for the BIOS test procedures and startup of
software to complete.
• Access the web interface via the IP address of the detector control unit or start using the SIMPLON API to
initialize and control the detector (see the API Reference documentation).
• For detectors with a CdTe sensor, it is necessary to wait up to 30 minutes after initializing the detector in
order to get the best data quality.
1
Some detectors do not require a dry air connection. Consult the Technical Specications for more information.
2
Some detectors do not require water cooling. Consult the Technical Specications for more information.
DECTRIS EIGER
®
2 User Manual v1.8.2 10 | 36