EasyManua.ls Logo

Delta DVP02LC-SL - Communication Settings

Delta DVP02LC-SL
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Load Cell Module DVP02LC-SL
DVP-PLC Operation Manual
21
The average values
Click on a value, and the enlarged characters of the value will appear.
The filter average values
Click on a value, and the enlarged characters of the value will appear.
Note: When the filtering function is not enabled, the window for the filter average value will display
“32767”. Once the filtering function is enabled, the average times have to be > 30, and the filter
percentage has to be set.
7. In the connection, if you would like to upload the data in DVP02LC-SL to software, click the icon
.
Or click the icon
if you would like to download all the parameters set in the software to
DVP02LC-SL.
6.2 Communication Settings
The communication setting window allows you to set up the communication formats for RS-232 and
RS-485, the characteristic value (eigenvalue) and measuring time. When all the settings are done, click
the icon
to download the parameters to DVP02LC-SL, or click “Upload” to display the parameters
of DVP02LC-SL in the software. Click “Default” and all the parameters set will return to the default
settings.