3 Unit description
3.2 Technical data
Dentsply Sirona
Operating Instructions Orthophos E
20
67 11 001 D3352
D3352.201.12.01.0203.2019
PAN sensor
Digital CCD line sensor, repluggable for panoramic exposure technique
Active sensor area, Pan
type:
138 mm x 6.48 mm
Detail resolution: 0.027 mm pixel size
Focus-sensor distance: 497 mm
Ceph sensor
Digital CCD line sensor, repluggable for panoramic or ceph exposure
technique
Active sensor area, Ceph
type:
230 mm x 6.48 mm
Detail resolution: 0.027 mm pixel size
Focus-sensor distance: 1,714 mm
3.2.2 Transport, storage, and operating conditions
Orthophos E
Transport and storage con-
ditions:
Temperature: -10°C – +70°C
(14°F – 158°F)
Relative humidity: 10% – 95%
+10
+40
95
10
106
70
kPa
Operating conditions: Ambient temperature: +10°C–+40°C
(50°F – 104°F)
Relative humidity: 10% - 95% (no con-
densation)
Operating altitude: ≤ 3,000m above sea level
Cephalometer
-40
+70
106
50
kPa
95
10
°C
Transport and storage con-
ditions:
Temperature: -40°C – +70°C (-40°F –
158°F)
Relative humidity: 10% – 95%
Air pressure: 500 hPa – 1060 hPa

Table of Contents

Other manuals for Dentsply Sirona Orthophos E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dentsply Sirona Orthophos E and is the answer not in the manual?

Dentsply Sirona Orthophos E Specifications

General IconGeneral
BrandDentsply Sirona
ModelOrthophos E
CategoryMedical Equipment
LanguageEnglish

Related product manuals