7 Malfunctions
7.3 Error description
67 11 001 D3352
D3352.201.12.01.0203.2019
99
E6 – Self-check
The error occurs spontaneously and without a corresponding operation.
1.
Acknowledge the error message.
Ä The error no longer occurs.
2.
If the error remains, switch the unit off and on again.
NOTE!
Observe waiting period!
Ä The error no longer occurs.
3.
If the error persists, contact your Customer Service.
E7 – Serious system error
The error occurs spontaneously and without a corresponding operation.
1.
Switch off the unit.
2.
Contact your Customer Service immediately.
Ä The unit is functional.
7.3.2 yy – Locality
The location may be a DX module number standing for an entire HW
function unit, or a logical SW function unit on board DX11 (central
control).
06
– Tube assembly
71
– Multipad user interface
10
– Central control DX 11; system hardware
11
– Central control DX 11; system software
12
– Central control DX 11; central CAN bus fault
13
– Central control DX 11; DX11, DX1 periphery (motor system of
stand, sensor system of stand)
14
– Central control DX 11; digital extension (HSI, network, etc.)
15
– Central control DX 11; configuration (wrong software, wrong
module constellation, etc.)
16
– Central control DX 11; Zero Management
20
– Central control DX 11; Framegrabber application
22
– Central control DX 11; 2D Imaging System
23
– Central control DX 11; 3D Imaging System
42
– Remote
61
– Diaphragm control
81
– Sensor
91
– Ceph digital