13.1.2 - ELECTROMAGNETIC IMMUNITY
Immunity Test
IEC
60601-1-2
Test Level
Compliance
Level
Electromagnetic
Environment Guidance
Electrostatic
Discharge
(ESD)
(IEC 61000-4-2)
±8kV
Contact
±15kV Air
±8kV
Contact
±15kV Air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Radiated RF
Electromagnetic
Field
(IEC 61000-4-3)
10 V/m
80 MHz to
2.7 GHz
3 V/m
2.7 to 6 GHz
10 V/m
80 MHz to
2.7 GHz
3 V/m
2.7 to 6 GHz
Minimum separation from
portable and mobile RF
communications equipment
should be more than:
E=6/d √P where
P = maximum power in W,
d = minimum separation
distance in m,
E = immunity level in V/m.
Electrical Fast
Transients/Burst
(IEC 61000-4-4)
±2 kV AC
Power Ports
±2 kV AC
Power Ports
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surges
(IEC 61000-4-5)
±0.5 kV,
±1 kV
Line-to-line
±0.5 kV,
±1 kV
Line-to-line
Mains power quality
should be that of a typical
commercial or hospital
environment.
Conducted
Disturbances
induced by
RFfields
(IEC 61000-4-6)
3 V
150 kHz to
80 MHz
6 V in
ISM bands
between
150 kHz and
30 MHz
80% AM at
1 kHz
3 V
150 kHz to
80 MHz
6 V in
ISM bands
between
150 kHz and
30 MHz
80% AM at
1 kHz
Minimum separation from
portable and mobile RF
communications equipment
should be more than:
E=6/d √P where
P = maximum power in W,
d = minimum separation
distance in m,
E = immunity level in V/m.
Power
Frequency
Magnetic Field
50/60Hz
(IEC 61000-4-8)
30 A/m 30 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage
Dips, Short
Interruptions
and Voltage
Variations
(
IEC 61000-4-11)
Various,
as specified
in Table 5
of IEC
60601-1-2:
2014
Various,
as specified
in Table 5
of IEC
60601-1-2:
2014
Mains power quality
should be that of a typical
commercial or hospital
environment.