Parameter
Description
Factory Value
P64 temperature below which the minimum temperature alarm is
activated during post-chill storage, with relation to P7, i.e. "P7
+ P64" (with reference to the cabinet probe)
0 = no alarm
0
P65 temperature above which the maximum temperature alarm is
activated during post-chill storage, with relation to P7, i.e. "P7
+ P65" (with reference to the cabinet probe)
0 = no alarm
0
P66 temperature below which the minimum temperature alarm is
activated during post-freezing storage, with relation to P8, i.e.
"P8 + P66" (with reference to the cabinet probe)
0 = no alarm
0
P67 temperature above which the maximum temperature alarm is
activated during post-freezing storage, with relation to P8, i.e.
"P8 + P67" (with reference to the cabinet probe)
0 = no alarm
0
P68 P64, P65, P66 and P67 differential 2
P69 storage operation start-up temperature alarm delay 0
P70 temperature alarm delay 0
P71 enable printing
1 = YES
0
P72 print interval 5
P73 HACCP alarm list deletion
1 = YES
0
P74 Baudrate
0=2400
1=4800
2=9600
3=19200
2
P75 Bit polarity
0=no polarity
1=even
2=odd
2
P76 Board address 1