SETUP AND CALIBRATION, Cont.
Weighing Units
With the display showing 8QLW press the PRINT key to show the current setting. If the value
displayed is acceptable, press the PRINT key again to save it. Otherwise use the UNITS key
to select the new setting and then press the PRINT key to save it. Allowable values are:
0 = Pounds & Ounces Only 1 = Kilograms Only
2 = Pounds & Ounces / Kilograms 3 = Kilograms / Pounds & Ounces
Scale Capacity
With the display showing &$3 press the PRINT key to show the current setting. Press the
UNITS key to enter the proper digit at the blinking location. Press the ZERO key to step to the
left and the next digit location. Repeat the process until all digits of the capacity have been
entered. After all digits have been correctly entered, press the PRINT key to store the capacity
and advance to the next step. Allowable values are 0 through 99.
Calibration
With the display showing &$/ press the PRINT key. The display will change to show the
current setting (0=NO). If the scale has been previously calibrated and you wish to skip
calibration and proceed to $&& , the Acceleration of Gravity, simply press the PRINT key and
the internal calibration factor will be retained.
To begin calibration, press UNITS to select (1=YES), then press the PRINT key. After
pressing the PRINT key the display will change to /2$G .
Load Calibration Weight
The scale will now display /2$G which is a prompt for the entry of the calibration weight value
and placement of this amount of test weights on the scale platform.
1. Make certain the scale platform is empty (remove any blankets or similar covering), then
place the calibrated test weights on the scale platform. NOTE! It is recommended that
a minimum of 50% of the scale’s capacity be used but 70% to 100% is preferred.
2. Press the PRINT key.
3. Determine the exact amount of test weights to be placed on the scale platform and
enter this value into the scale by using the UNITS and PRINT keys in the same manner
used to enter the scale’s capacity.
4. Verify that the numbers entered are the same as the total weight of test weights, and
the least significant digit agrees with the scale interval.
5. Press the PRINT key.
After a moment the scale will display the message XQ/2$G which is a request that the test
weights be removed from the scale platform. Remove the weights then press the PRINT key.
The calculated calibration factor is now stored in the scale’s nonvolatile memory.