13
SETUP AND CALIBRATION, Cont.
USA (Domestic or International)
The display will change to show 86$ . This is the prompt to select whether the scale is used
in the USA (Domestic) or outside the US (International). Press the PRINT key to show current
setting. If the value displayed is acceptable, press the PRINT key again to save it. Otherwise
use the UNITS key to select the new setting and then press the PRINT key to save it.
USA = 1 (Domestic) USA = 0 (International)
No Zero Limit +/- 2% Zero Limit
Threshold Lock Feature Enabled Threshold Lock Feature Disabled
Lamp test on power up enabled (display
segments 1 second on, 1 second off)
Digital Filter Level Selection
Your scale will arrive with the factory filter setting (1=minimal) already entered. Please check
with your scale service technician should you wish to change the programmed filter level and
break range.
With the display showing )/W , press the PRINT key to show the current setting. If the value
displayed is acceptable, press the PRINT key again to save it. Otherwise use the UNITS key
to select the new setting and then press the PRINT key to save it. Four levels of filtering are
available. They are as follows:
0 = NO FILTERING 1 = MINIMAL FILTERING
2 = MODERATE FILTERING 3 = CUSTOM FILTERING
NOTE: Selection 3, Custom Filtering is used when 0, 1 or 2 are inadequate.
) - Filter Level
If you select Custom Filtering, the scale will display ) . Press the PRINT key to show the
current setting for the Filter Level. The filter level is a number from 1 to 16 that
corresponds to the level of filtering with 16 being the greatest filtering and 1 the least.
Use the UNITS and ZERO keys to select the filter level, and then press the PRINT key to
save the setting. Allowable values are: 1 through 16.
EU - Break Range
Next, the scale will display EU . Press the PRINT key to show the current setting for the
Break Range. The break range is a number from 1 to 64 that corresponds to the number
of division change to break out of filtering. Use the UNITS and ZERO keys to select the
break range value, and then press the PRINT key to save the setting. Allowable values
are: 1 through 64.
Sample Rate
The display will change to show 6U . Press the PRINT key to show the current setting. The
sample rate may be set from a minimum of 1 sample per second to a maximum of 10 samples
per second in one sample per second intervals. Use the UNITS and ZERO keys to select the
desired sample rate and then press the PRINT key to save the setting. Allowable values are:
1 through 10.