Programming Manual
PC 2003 Amplifier
English
Page 14/14
PROPC2003EN – Index A
SENSOR PARAMETERS
SERIAL PORT
>PROGRAM SETTINGS
ADMIN
CONFIGURATION
GOOD MEASUREMENT
>MI/MA MEASUREMENT
Memorization of the outputs min. and max.:
STORED: When a part is bad min. or max., the
outputs close (or open if Normally Closed) and stay
close until the next valid signal. The state of the
output is stored. (adjustment by default)
NOT STORED: When a part is bad, the outputs
min. or max. close 50ms minimum and reopen.
The state of the outputs is not stored.
>LOCK
UNLOCK
CODE CHANGE
FACTORY VALUES
This function allows locking all the adjustments
(preload, reference, tolerances, etc) made, leaving
only the display to the user. The original code is
0000.
LOCK
>UNLOCK
CODE CHANGE
FACTORY VALUES
This function allows unlocking all the adjustments
(preload, reference, tolerances, etc.).
LOCK
UNLOCK
>CODE CHANGE
FACTORY VALUES
Change the original code (0000) by the code you
want.
MEASURE
OPTIONS
>PARAMETERS
DISPLAYS
SENSOR PARAMETERS
SERIAL PORT
PROGRAM SETTINGS
>ADMIN
LOCK
UNLOCK
CODE CHANGE
>FACTORY VALUES
Return to the box’s original parameter.
Warning: This will force you to redo the preload,
referencing and adjustment of the tolerances.
>BRIGHT./CONTRAST
LANGUAGE
MEASUREMENT UNIT
Adjust the brightness of the LCD screen. Minimum 0 and maximum 15.
NB : After the factory values are returned to, the value will be 8
Adjust the contrast of the LCD screen. Minimum 0 and maximum 15.
NB : After the factory values are returned to, the value will be 8
BRIGHT./CONTRAST
>LANGUAGE
MEASUREMENT UNIT
Select the language, Français, English, Italiano, Espagnol and Deutsch
NB : After the factory values are returned to, the value will be French.
PART SETUP
OPTIONS
PARAMETERS
>DISPLAYS
BRIGHT./CONTRAST
LANGUAGE
>MEASUREMENT UNIT
Select the measurement unit, Millimeter or Inch.
NB : After the factory values are returned to, the value will be MILLIMETER
In mm: 3 digits are displayed before the decimal point + 3 digits after the
decimal point.
Maximum 999.999mm
In Inch: 2 digits are displayed before the decimal point + 4 digits after the
decimal point.
Maximum 21.4747in
: Screens appear only if you have a PC2003 DUAL CHANNEL device.