0044
EN
DeVilbiss® 5-Liter Oxygen Concentrator
Instruction Guide
WARNING–Read instruction guide before operating this equipment.
ASSEMBLED IN USA
NO SMOKING
ES
Guía de instrucciones del concentrador de
oxígeno de 5-litros de DeVilbiss®
ADVERTENCIA–Lea la guía de instrucciones antes de poner a funcionar este equipo.
ENSAMBLADO EN EE. UU.
NO FUMAR
FR
Guide d’instructions du concentrateur
d’oxygène 5-litres DeVilbiss®
AVERTISSEMENT–Lire le mode d’emploi avant d’utiliser ce dispositif.
ASSEMBLÉ AUX ÉTATS-UNIS
NE PAS FUMER
DE
DeVilbiss® 5-Liter-Sauerstoffkonzentrator
Bedienungsanleitung
WARNUNG–Vor Inbetriebnahme des Gerätes Bedienungshinweise lesen.
HERGESTELLT IN DEN USA
RAUCHEN VERBOTEN
IT
Concentratore di ossigeno da 5-litri
DeVilbiss® Istruzioni per l’uso
AVVERTENZA–Leggere il manuale di istruzioni prima di usarel’apparecchio
ASSEMBLATO NEGLI U.S.A.
NON FUMARE
NL
Instructiehandleiding DeVilbiss® 5-liter
zuurstofconcentrator
WAARSCHUWING–Lees dit instructiehandboekje zorgvuldig door voordat u het
apparaat gaat gebruiken.
GEMONTEERD IN DE VERENIGDE STATEN
VERBODEN TE ROKEN
TR
DeVilbiss® 5-Litre Oksijen Konsantratörü
Kullanım Kılavuzu
UYARI–Cihazı kullanmaya başlamadan önce bu kılavuzu okuyunuz.
ABD’DE MONTE EDILMIŞTIR
SİGARA İÇİLMEZ
PT
Manual de instruções do Concentrador de
oxigênio DeVilbiss® de 5-litros
ADVERTÊNCIA– Leia o manual de instruções antes de operar este equipamento.
MONTADOS NOS EUA
PROIBIDO FUMAR
PL
Instrukcja obsługi 5-litrowego
koncentratora tlenu DeVilbiss®
OSTRZEŻENIE – LPrzeczytaj instrukcję obsługi przed rozpoczęciem korzystania z tego
urządzenia.
ZMONTOWANO W STANACH ZJEDNOCZONYCH
NIE PALIĆ

AR
5®DeVilbiss





525 Series
Serie de 525
Série 525
525 Serie
Serie da 525
525 Serie
525 Serisi
Série 525
Seria 525
525 

Table of Contents

Questions and Answers

  • A
    austin97Aug 1, 2025
    What to do if DeVilbiss 525KS-LT unit does not operate?
    • P
      Paul BowmanAug 1, 2025
      If the DeVilbiss Medical Equipment unit isn't operating, all lights are off, and an audible alert is pulsing, first ensure the power cord is properly inserted into the wall outlet. For 230-volt units, check the mains connection on the back. If it still doesn't work, check your home circuit breaker and reset it if needed. If problems persist, try a different wall outlet. Also, check the oxygen concentrator circuit breaker and reset it if necessary by pressing the reset button located below the power switch. If these steps don't resolve the issue, contact your DeVilbiss provider.
  • R
    rodriguezdavidAug 5, 2025
    What to do if red service required light is on DeVilbiss 525KS-LT?
    • T
      Tyler JenkinsAug 5, 2025
      If the red Service Required light is illuminated on your DeVilbiss Medical Equipment and an audible alert is sounding, check the following: * **Air filter:** If it's dirty, clean it according to the instructions. * **Exhaust area:** Ensure nothing is blocking the unit's exhaust. * **Cannula, mask, or tubing:** Detach these to see if proper flow returns. If it does, clean or replace the blocked component. * **Humidifier bottle:** Detach to check if flow improves; clean or replace if needed. * **Flow meter:** Ensure it's set to the prescribed rate. If issues persist, contact your DeVilbiss provider.
  • J
    James LawsonAug 10, 2025
    Why is the yellow low oxygen light on DeVilbiss 525KS-LT Medical Equipment?
    • J
      John EllisonAug 10, 2025
      If the yellow Low Oxygen light is on your DeVilbiss Medical Equipment, allow the unit up to 15 minutes to complete its start-up period.
  • T
    Tamara LoganAug 13, 2025
    What to do if yellow low oxygen light is on and intermittent audible signal is sounding on DeVilbiss Medical Equipment?
    • E
      Edward SimsAug 14, 2025
      If the yellow Low Oxygen light is on and an intermittent audible signal is sounding on your DeVilbiss Medical Equipment, make sure the flow meter is properly set to the prescribed number. Note that the maximum flow meter setting is 3 LPM when filling an oxygen bottle from the auxiliary port. Also, check the air filter and clean it if it's dirty. Finally, check the exhaust area to ensure nothing is blocking the unit's exhaust. If these steps don't resolve the issue, contact your DeVilbiss provider.
  • P
    Pamela SmithAug 18, 2025
    What to do if red service required light is on and intermittent audible signal is sounding on DeVilbiss 525KS-LT?
    • C
      Crystal HarrisonAug 18, 2025
      If the red Service Required light is on and an intermittent audible signal is sounding on your DeVilbiss Medical Equipment, ensure the flow meter is properly set to the prescribed number, noting that the maximum flow meter setting is 3 LPM when filling an oxygen bottle from the auxiliary port. Check the air filter and clean it if needed. Also, check the exhaust area to ensure it's not blocked. If these steps don't resolve the issue, or if you suspect an Electronic Assembly Malfunction, turn the unit off, switch to your reserve oxygen system, and contact your DeVilbiss provider immediately.
  • T
    Toni GarciaAug 22, 2025
    What to do if DeVilbiss Medical Equipment makes audible low-frequency vibration sound?
    • J
      Jamie HernandezAug 22, 2025
      If you detect an audible low-frequency vibration sound from your DeVilbiss Medical Equipment, turn the unit off, switch to your reserve oxygen system, and contact your DeVilbiss provider immediately, as this may indicate an Electronic Assembly Malfunction.
  • C
    Christian GarciaAug 27, 2025
    What to do if any other problems occur with DeVilbiss oxygen concentrator?
    • B
      Brandon GonzalezAug 27, 2025
      Turn your unit “Off.” Switch to your reserve oxygen system and contact your DeVilbiss provider immediately.

Summary

Important Safeguards

Introduction

Essential Performance

Highlights the device's primary function of delivering continuous oxygen-enriched gas with alarms for failures.

Important Parts of Your Concentrator

Setting Up Your Oxygen Concentrator

Oxygen Tubing Connection

Instructions for attaching oxygen tubing to the concentrator's oxygen outlet.

Oxygen Tubing Connection With Humidification

Steps for connecting oxygen tubing when a humidifier bottle is used with the concentrator.

Operating Your DeVilbiss Oxygen Concentrator

DeVilbiss OSD Operation

Caring For Your DeVilbiss Oxygen Concentrator

Air Filter Maintenance

Instructions for inspecting, cleaning, and replacing the air filter for the oxygen concentrator.

Troubleshooting

Minor Troubleshooting Chart

A chart detailing common symptoms, possible causes, and remedies for concentrator malfunctions.

Overview of Alarms

Alarm Conditions

Details the meaning of visual and audible alarms for start-up, low oxygen, and malfunction.

Return and Disposal

Cleaning and Disinfection Between Patients

Procedures for cleaning and disinfecting the concentrator between patient uses.