DeVilbiss® 5-Liter Oxygen Concentrator
Instruction Guide
WARNING–Read instruction guide before operating this equipment.
ASSEMBLED IN USA
NO SMOKING
es
Guía de instrucciones del concentrador de
oxígeno de 5-litros de DeVilbiss®
ADVERTENCIA–Lea la guía de instrucciones antes de poner a funcionar este equipo.
ENSAMBLADO EN EE. UU.
NO FUMAR
fr
Guide d’instructions du concentrateur
d’oxygène 5-litres DeVilbiss®
AVERTISSEMENT–Lire le mode d’emploi avant d’utiliser ce dispositif.
ASSEMBLÉ AUX ÉTATS-UNIS
NE PAS FUMER
de
DeVilbiss® 5-Liter-Sauerstoffkonzentrator
Bedienungsanleitung
WARNUNGVor Inbetriebnahme des Gerätes Bedienungshinweise lesen.
HERGESTELLT IN DEN USA
RAUCHEN VERBOTEN
it
Concentratore di ossigeno da 5-litri
DeVilbiss® Istruzioni per l’uso
AVVERTENZA–Leggere il manuale di istruzioni prima di usare l’apparecchio
ASSEMBLATO NEGLI U.S.A.
NON FUMARE
nl
Instructiehandleiding DeVilbiss® 5-liter
zuurstofconcentrator
WAARSCHUWING–Lees dit instructiehandboekje zorgvuldig door voordat u het
apparaat gaat gebruiken.
GEMONTEERD IN DE VERENIGDE STATEN
VERBODEN TE ROKEN
tr
DeVilbiss® 5-Litre Oksijen Konsantratörü
Kullanım Kılavuzu
UYARI–Cihazı kullanmaya başlamadan önce bu kılavuzu okuyunuz.
ABD’DE MONTE EDILMIŞTIR
SİGARA İÇİLMEZ
pt
Manual de instruções do Concentrador de
oxigênio DeVilbiss® de 5-litros
ADVERTÊNCIA– Leia o manual de instruções antes de operar este equipamento.
MONTADOS NOS EUA
PROIBIDO FUMAR
pl
Instrukcja obsługi 5-litrowego
koncentratora tlenu DeVilbiss®
OSTRZEŻENIE – Przeczytaj instrukcję obsługi przed rozpoczęciem korzystania z tego
urządzenia.
ZMONTOWANO W STANACH ZJEDNOCZONYCH
ZAKAZ PALENIA

ar
5DeVilbiss®



0044
525 Series
Serie de 525
Série 525
525 Serie
Serie da 525
525 Serie
525 Serisi
Série 525
Seria 525


Questions and Answers

  • T
    taylor17Sep 9, 2025
    What to do if I have any other problems with my DeVilbiss Oxygen Equipment?
    • E
      emily89Sep 9, 2025
      Turn your unit off and switch to your reserve oxygen system immediately. Contact your DeVilbiss provider for assistance.
  • C
    Carla ClementsAug 23, 2025
    Why is the red service required light on my DeVilbiss Oxygen Equipment illuminated?
    • M
      Michael SalazarAug 23, 2025
      The red Service Required light may be on for several reasons. Check the air filter and clean it if it's dirty, following the cleaning instructions. Also, ensure nothing is blocking the exhaust area of the unit. Check the cannula, face mask, or oxygen tubing for blockages or defects; detach them to see if proper flow is restored, and clean or replace if needed. Inspect the humidifier bottle for blockages or defects, detaching and cleaning or replacing it if necessary. Finally, ensure the flow meter is set to the prescribed flow rate. If problems continue, contact your DeVilbiss provider.
  • M
    Margaret WalkerAug 25, 2025
    Why is the yellow low oxygen light on my DeVilbiss Oxygen Equipment?
    • B
      benjaminhartAug 25, 2025
      The yellow Low Oxygen light may be triggered if the flow meter isn't properly set; ensure it's set to the prescribed number. It could also indicate a blocked air filter, so check and clean it if necessary, following the cleaning instructions. Another possibility is a blocked exhaust area; make sure nothing is restricting the unit's exhaust. If the issue continues, contact your DeVilbiss provider.
  • B
    Brian DeckerAug 26, 2025
    What to do if the red service required light is on and an intermittent audible signal is sounding on my DeVilbiss Oxygen Equipment?
    • A
      Amanda WatsonAug 27, 2025
      Make sure that the flow meter is properly set to the prescribed number. Check the air filter and clean it if it is dirty, following the cleaning instructions. Check the exhaust area and ensure that nothing is restricting the unit exhaust. If the above remedies do not work, contact your DeVilbiss provider.
  • M
    maxwestAug 27, 2025
    What to do if the red service required light is on for DeVilbiss 525?
    • A
      amywilsonAug 27, 2025
      If the DeVilbiss Medical Equipment shows a red 'Service Required' light accompanied by an intermittent audible signal, first, ensure the flow meter is correctly set to the prescribed number. Then, check the air filter and clean it if it's dirty, following the cleaning instructions. Finally, inspect the exhaust area to ensure nothing is blocking the unit's exhaust.
  • S
    shuntAug 29, 2025
    Why is the yellow low oxygen light on for DeVilbiss 525 Medical Equipment?
    • Z
      Zachary NortonAug 29, 2025
      If your DeVilbiss Medical Equipment displays a yellow 'Low Oxygen' light, possibly accompanied by an intermittent audible signal, start by verifying that the flow meter is accurately set to the prescribed level. Next, examine the air filter and clean it according to the provided cleaning instructions if it appears dirty. Lastly, inspect the exhaust area to confirm that nothing is obstructing the unit's exhaust.
  • J
    Joshua LloydSep 18, 2025
    Why does my DeVilbiss Medical Equipment make a low-frequency vibration sound?
    • D
      diana58Sep 18, 2025
      If your DeVilbiss Medical Equipment is making an audible low-frequency vibration sound, it could be due to an Electronic Assembly Malfunction. In this case, turn your unit off, switch to your reserve oxygen system, and contact your DeVilbiss provider immediately.
  • J
    Jared SotoSep 22, 2025
    What should I do if any other problems occur with my DeVilbiss oxygen concentrator?
    • R
      Richard JacksonSep 22, 2025
      Turn your unit “Off.” Switch to your reserve oxygen system and contact your DeVilbiss provider immediately.
  • J
    Jeremy JensenSep 7, 2025
    What does it mean if both the green normal oxygen and yellow low oxygen lights are on or off on my DeVilbiss Oxygen Equipment?
    • A
      Amber PaceSep 8, 2025
      This indicates an OSD malfunction. Contact your DeVilbiss provider for assistance.
  • A
    Austin KleinSep 4, 2025
    What should I do if my DeVilbiss 525 Medical Equipment is vibrating?
    • C
      Cody NixonSep 5, 2025
      If you detect an audible low-frequency vibration sound from your DeVilbiss Medical Equipment, turn the unit off immediately. Switch to your reserve oxygen system and contact your DeVilbiss provider right away.

Summary

Symbol Definitions

General Warning Symbol

Symbol used to indicate hazardous situations to avoid.

Important Information Symbol

Symbol used to indicate important information you should know.

Note and Information Symbol

Symbol used to indicate notes, useful tips, recommendations, and information.

Important Safeguards

Introduction

Intended Use

Specifies the purpose of the DeVilbiss 5 Liter Oxygen Concentrator.

Indications For Use

Details when the DeVilbiss Oxygen Concentrator should be used.

Contraindications

Lists situations or patients for whom the device is not suitable.

Essential Performance

Describes the primary function and alarm indications of the concentrator.

Service Life

Explains the expected operational lifespan and factors affecting it.

Why Physician Prescribed Supplemental Oxygen

Explains the benefits of supplemental oxygen therapy for patients.

How Your Oxygen Concentrator Works

Details the electrical operation and oxygen separation process.

Important Parts of Your Concentrator

Front View (Figure A)

Identifies and describes the controls and indicators on the front of the unit.

Back View (Figure B)

Identifies and describes the ports and vents on the back of the unit.

Accessories

Lists and describes compatible accessories like humidifiers and tubing.

Setting Up Your Oxygen Concentrator

Oxygen Tubing Connection

Instructions for attaching oxygen tubing to the concentrator outlet.

Oxygen Tubing Connection With Humidification

Steps for connecting oxygen tubing and humidifier bottle.

Operating Your Oxygen Concentrator

DeVilbiss OSD Operation

Explains the Oxygen Sensing Device (OSD) and its indicator lights.

Reserve Oxygen System

Describes the reserve oxygen system and its use in case of power loss or failure.

Caring for Your Oxygen Concentrator

Cleaning Accessories

Instructions for cleaning and replacing accessories like cannula, tubing, and humidifier.

Cleaning Filter Door Vents

Guidance on cleaning the filter door vents to remove dust.

Cleaning Exterior Cabinet

Weekly cleaning instructions for the concentrator's outer cabinet.

Troubleshooting

Overview of Alarms and Service Indicators

Details the device's alarm system, LED indicators, and required actions.

Specifications

Electromagnetic Compatibility Information

Warranty

Warranty Claims Submissions

Procedure for submitting warranty claims and returning the unit.

Ordering and Returning Parts

Customer Service Contact Information

Provides contact details for DeVilbiss customer service and international department.

Ordering Non-Warranty Replacement Parts

Instructions for ordering parts not covered under warranty.

Return and Disposal

Providers Notes - Cleaning and Disinfection

Service and Maintenance Guidance

Initial Inspection

Steps for inspecting the unit upon receiving it and after setup.

Oxygen Provider Preventive Maintenance Guidance

Scheduled maintenance and service tasks for oxygen providers.

DeVilbiss 525 Series Specifications

General IconGeneral
TypeOxygen Concentrator
Flow Rate0.5 to 5 LPM
Outlet Pressure5.5 PSI ± 0.5 PSI
Operating Temperature41°F to 104°F (5°C to 40°C)
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Operating Humidity15% to 95% non-condensing
Oxygen Concentration93% ±3%
Sound Level48 dBA
Power Consumption310 Watts
Weight36 lbs
Altitude0 to 4000 m (0 to 13, 123 ft)
AlarmsPower Failure, High/Low Pressure, Low Oxygen