EasyManua.ls Logo

DeVilbiss 525 Series - DeVilbiss OSD Operation; Reserve Oxygen System; Caring for Your Oxygen Concentrator; Cleaning Accessories

DeVilbiss 525 Series
14 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN - 9
DANGER
Oxygen makes it easier for a re to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is
turned ON but not in use, the oxygen will make the materials ammable. Turn the oxygen concentrator OFF when not in use to prevent oxygen enrichment.
NOTE If the audible signal alarms but the unit is not operating, there is no power to the unit. Refer to the Troubleshooting chart on page 10, and contact your
DeVilbiss provider if necessary.
NOTE If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Troubleshooting chart on page 10, and contact your
DeVilbiss provider if necessary.
2. Check the ow meter to make sure that the ow meter ball is centered on the line next to the prescribed number of your ow rate.
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the ow of oxygen – consult your physician.
NOTE Your DeVilbiss provider may have preset the ow meter so that it cannot be adjusted.
NOTE If the ow meter knob is turned clockwise, the ow decreases (and eventually will shut off the oxygen ow). If the knob is turned counterclockwise, the ow
increases.
NOTE For prescriptions of 5 LPM, be sure the ball is centered on the 5 liter line. The ball should not touch the red line. Setting the
ow higher than 5 may cause the oxygen purity level to drop.
NOTE The low-ow alarm may activate if the ow meter ball is set at or below 0.2 lpm. The unit will continue to run; however, the
Service Required light will come on accompanied by an audible alarm. Adjust the ow meter to your prescribed ow.
3. Your DeVilbiss concentrator is now ready for use, properly position the cannula with the nasal prongs facing upward, insert the prongs
into nose. Wrap the cannula tubing over the ears and position in front of body (Figure 4). Allow 15 minutes for the oxygen concentrator
to reach stated performance.
DeVilbiss OSD
®
Operation
The OSD (Oxygen Sensing Device) is a device within your concentrator that monitors the oxygen produced by your unit.
The OSD lights on the top panel are dened as follows:
Green Normal Oxygen light–acceptable oxygen level.
Yellow Low Oxygen light– below an acceptable oxygen level.
If the oxygen purity falls below the acceptable level: The green Normal Oxygen light will shut off, the yellow Low Oxygen light will illuminate, and an intermittent audible signal
will sound.
Refer to the Troubleshooting section in this guide on page 10, and switch to your reserve oxygen system. Do not attempt any other maintenance. Contact your DeVilbiss
provider immediately.
RESERVE OXYGEN SYSTEM
As a precaution, your DeVilbiss provider may supply you with a reserve oxygen system. If your unit loses electrical power or fails to operate correctly, the Patient Alert System
will sound to signal you to switch to your reserve oxygen system (if provided) and contact your DeVilbiss provider. Please contact your oxygen provider if you have questions
regarding a reserve oxygen system.
CARING FOR YOUR DEVILBISS OXYGEN CONCENTRATOR
DeVilbiss recommends using only original DeVilbiss parts and lters in order to guarantee reliable operation of the product.
WARNING
Do NOT use lubricants, oils or grease.
Before attempting any cleaning procedures, turn the unit “OFF.”
Cannula/Mask, Tubing and Humidifier Bottle
Clean and replace the cannula/mask, tubing and humidier bottle according to the manufacturer’s instructions.
Filter Door with Vents
Inspect the vents periodically, and wipe with a dry cloth as needed to remove dust.
Exterior Cabinet
Clean the concentrator exterior cabinet weekly by using a damp cloth and wiping dry; the vents can also be wiped with a damp cloth.
Cleaning
Recommended cleaning interval Number of cleaning cycles * Compatible cleaning method
Outer Cabinet 7 days 260 Water, use only a damp cloth
Filter Door Vents 7 days 260 Wipe with dry cloth, or a cloth dampened with water
to remove dust.
Oxygen Outlet Connector 7 days 104 Mild dish soap (2 tbsp) and warm water (2 cups)
* number of cleaning cycles determined by recommended cleaning interval and expected service life
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken. Do not remove the concentrator cabinet. The cabinet
should only be removed by a qualied DeVilbiss technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or cleaning agents.
CAUTION
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid
inadvertent damage.
4
SE-525KS-1

Related product manuals