DeVilbiss 525KS - Service Manual

DeVilbiss 525KS
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
For Product
Numbers:
525DS
525KS
SERVICE MANUAL
DANGER NO SMOKING
CAUTION-Federal (U.S.A.) law restricts this device
to sale by or on the order of a physician.
DeVi l b i s s
®
5-li t e r Co m p a C t ox y g e n Co n C e n t r a t o r

Other manuals for DeVilbiss 525KS

Questions and Answers

  • G
    Gene BrownSep 8, 2025
    What to do if the yellow Low Oxygen light is on for DeVilbiss Health Care Products?
    • G
      Geoffrey MedinaSep 8, 2025
      If the yellow Low Oxygen light is on, or the yellow Low Oxygen light is on and the intermittent audible signal is sounding on your DeVilbiss Health Care Products unit, make sure the flow meter is properly set to the prescribed number. Check the air filter and wash it if it's dirty. Also, verify that nothing is blocking the exhaust area. If these steps don't solve the problem, contact your DeVilbiss provider.
  • E
    Elizabeth MoralesSep 12, 2025
    What to do if the Service Required light is on for DeVilbiss Health Care Products?
    • G
      Gary MurraySep 12, 2025
      If your DeVilbiss Health Care Products unit operates, the Power light is on, but the Red Service Required light is illuminated, and an audible alert may be sounding, consider the following: Check the air filter and wash it if it is dirty. Check the exhaust area to ensure nothing is restricting it. Detach the cannula or face mask; if proper flow is restored, clean or replace them. Disconnect the oxygen tubing at the oxygen outlet; if flow returns, inspect the tubing for obstructions or kinks and replace if needed. Detach the humidifier; if proper flow is obtained, clean or replace it. Finally, ensure the flow meter is set to the prescribed rate. If problems persist, contact your DeVilbiss provider.
  • M
    marie19Sep 12, 2025
    What to do if other problems occur with my DeVilbiss Health Care Products?
    • W
      William RamirezSep 13, 2025
      If any other problems occur with your DeVilbiss Health Care Products oxygen concentrator, turn the unit off, switch to your reserve oxygen system, and contact your DeVilbiss provider immediately.
  • M
    Marc ComptonSep 14, 2025
    Why are both oxygen lights on/off on my DeVilbiss Health Care Products?
    • S
      shayesSep 14, 2025
      If both the green Normal Oxygen and the yellow Low Oxygen lights are either on or off on your DeVilbiss Health Care Products unit, this may indicate an OSD malfunction. Contact your DeVilbiss provider for assistance.
  • D
    David BennettSep 17, 2025
    What to do if I hear a vibration sound from my DeVilbiss Health Care Products?
    • T
      trevorperezSep 17, 2025
      If you detect an audible low-frequency vibration sound from your DeVilbiss Health Care Products unit, turn it off. Switch to your reserve oxygen system and immediately contact your DeVilbiss provider.

Summary

GENERAL INFORMATION

UNPACKING AND SETUP

Operating Instructions and Patient Alerts

Details how to operate the unit and explains the function of the patient alert system for monitoring.

MAINTENANCE

Preventative Maintenance Schedule

Details periodic maintenance schedules and tasks for homecare providers to ensure unit performance.

TROUBLESHOOTING

Troubleshooting Charts

Comprehensive charts detailing specific issues, possible causes, and remedies for unit malfunctions.

COMPONENT TESTING, REPAIR, AND REPLACEMENT

Compressor and Related Components

Details testing, replacement, and maintenance of the compressor, capacitor, and cooling fan.

SPECIFICATIONS

DeVilbiss 525KS Specifications

General IconGeneral
Flow Rate0.5 to 5 L/min
Oxygen Concentration93% ± 3%
Sound Level48 dBA
Weight16.3 kg
Dimensions62.2 cm x 34.2 cm x 30.4 cm
Operating Temperature10°C to 40°C (50°F to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
AlarmsPower failure, Low Oxygen
Altitudeup to 2286 m