Specifications
17
Model 9001 Pressure swings (cmH
2
O peak to peak at 500mL tidal volume)
Breaths per minute 10 15 20
@ 7 cmH
2
O 1.0 1.0 1.1
@ 13 cmH
2
O 1.0 1.0 1.1
@ 20 cmH
2
O 1.1 1.1 1.2
Warranty.................................................................................Two-year limited
Equipment classification with respect to protection from electric shock........Class II
Degree of protection from electric shock................................Type BF Applied Part
Degree of protection against ingress of liquids.............................................None
Equipment not suitable for use in the presence of a flammable anaesthetic mixture
with air or with oxygen or nitrous oxide.
Mode of operation ...........................................................................Continuous
Definitions
Display – purpose, type, information displayed. The LCD display shows the following
information in each model:
Model 9000 Model 9001 Description
Set Pres Set Pres CPAP pressure prescribed for the patient in cmH
2
O
Altitude Altitude setting one of four ranges
Fine Adj Fine adjustment for fine tuning pressure using an
external manometer
Dly Time Dly Time Delay time prescribed for the patient in minutes
Dly Pres Dly Pres Delay pressure prescribed for the patient in cmH
2
O
Compl Hrs Compl Hrs Compliance meter, shows accumulated time while
breathing
Unit Hrs Unit Hrs Unit hour meter, shows accumulated time of blower
operation
Firmware Firmware Firmware version installed in the CPAP
Serial # Serial # Serial number of the CPAP
Language Language Language selection for LCD display messages
Off Off Blower is OFF AC or DC power is connected
Error Error An error has been detected by the firmware, the
CPAP is in a fail safe mode
A-900
Specifications
16
SPECIFICATIONS
Technical Specifications
Size .............................................................................4.0"H x 7.5"W x 10.8"D
(10.2 cm x 19.6 cm x 26.2 cm)
Weight....................................................................................3.6 lbs. (1.7 kg)
Electrical Requirements (9000/9001) ...................................100-240V~, 50/60 Hz
9001 Only........................................................................12 VDC @ 3 A max
Maximum Power Consumption...............................................................65 watts
CPAP Pressure Range......................................................................3-20 cm H
2
O
Pressure Delay Time..................................................0, 10, 20, 30 or 45 minutes
Delay Pressure Range........................................3 cm H
2
O to prescription pressure
Operating Temperature Range ...................................41°F to 104°F (5°C to 40°C)
Operating Humidity Range ....................................5% to 95% RH non-condensing
Operating Atmospheric Conditions .....................................Sea level to 9,000 feet
Storage & Transportation Temperature Range. .........-4°F to 140°F (-20°C to +60°C)
Storage & Transportation Humidity Range................5% to 95% RH non-condensing
Maximum Limited Pressure .................................................................28 cmH
2
O
Sound Level........................................30 dBA (@ 10 cmH
2
O, 1 meter from CPAP)
Filter specs ..................................................Standard Filter > 3 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Max flow rates Model 9000 Model 9001
83 L/m 73 L/m @ 7 cmH
2
O (1/3 max pressure)
134 L/m 118 L/m @ 13 cmH
2
O (2/3 max pressure)
163 L/m 164 L/m @ 20 cmH
2
O (max pressure)
Pressure Stability
Between maintenance intervals Model 9000 = ± 2 cmH
2
O
(maintenance interval = 1 year) Model 9001 = ± 1 cmH
2
O
Model 9000 Pressure swings (cmH
2
O peak to peak at 500mL tidal volume)
Breaths per minute 10 15 20
@ 7 cmH
2
O 1.0 1.0 1.0
@ 13 cmH
2
O 1.0 1.0 1.0
@ 20 cmH
2
O 1.3 1.3 1.4
A-900