SECTION
6
TROUBLESHOOTING
GUIDE
Symptom
o.
Possible
Problems
Suggested
Remed
Visible
|
Audible
|
Compressor
Other
99
y
Alarm
Alarm
On
or
Off
Symptoms
1.
OFF
ON
OFF
On/Ott
Line
cord
not
properly
installed
|
Insert
plug
in
receptacle
or
replace
switch
not
[|
or
defective
line
cord
illuminated
|
yo
power
at
receptacle
Check
building
circuit
breaker
or
fuse,
or
have
house
wiring
checked
by
qualified
electrician.
Circuit
may
be
fully
loaded
with
other
appliances
and
another
receptacle
may
be
required.
Line
cord
disconnected
at
Reconnect
line
cord
at
terminal
termina!
block block
Defective
On/Off
switch
Replace
switch
Blown
fuse
in
power
connector
Replace
fuse
2.
ON
ON
OFF
On/Off
Defective
MC
board
Replace
MC
board
İN
ted
Loose
Compressor
Wire
Tighten
or
attach
wire
utumına
.
Defective
compressor
Replace
compressor
Defective
Capacitor
Replace
Capacitor
Upper
and
lower
wiring
harness
|
Reconnect
wiring
harness
disconnected
3.
ON
ON
OFF
Unit
warm
Compressor
overheated
due
to:
o
the
a
Occluded
filters
Replace
filters
U
an
.
. : :
cannot
be
Restricted
input
or
output
air
Remove
obstruction
restarted
passage
for
several
|
Low
line
voltage
Check
line
voltage,
use
alternate
minutes
circuit
independent
of
other
appliances
Defective
cooling
fan
Replace
cooling
fan
Defective
compressor
Replace
compressor
4.
OFF
OFF
ON
Pulsating
Felt
and/or
intake
bacteria
Check
filters
and
replace
if
air
noise
filters
not
in
place
or
defective
necessary
Compressor
intake
hose
Reconnect
hose
disconnected
5.
OFF
OFF
ON
Excessive
Defective
or
loose
motor
Replace
or
tighten
motor
mounts
noise
mounts
Defective
compressor
Replace
compressor
Defective
cooling
fan
Replace
cooling
fan