physician prescribed oxygen for your sleep therapy, you can add the oxygen in one of two ways:
1. Obtain the optional oxygen adapter (part #7353D-601) from your homecare provider and connect it to the air supply
port on the IntelliPAP. Connect one end of the air supply tubing to the open end of the adapter and the other end of the
air supply tubing to the mask.
2. If your mask is equipped with an oxygen port, your physician may request that you apply oxygen directly to your mask.
Air-inlet lter (4/pk) DV51D-602 Heater connection cover DV51D-605
Fine particle lter (4/pk) DV51D-603 Oxygen adapter 7353D-601
Air supply tubing (6’) DV51D-629 Carrying case DV51D-610
Air supply plug DV51D-604 Heated Humidier DV5HH
Humidier Chamber DV5C Chamber Sealing Gasket DV5C-614
AC power cord (USA) DV51D-606 AC power cord (Australia) DV51D-609
AC power cord (UK) DV51D-608 DC power cord DV51D-619
AC power cord (EU) DV51D-607
DC battery clamp-on adapter (requires DC Power
Cord DV51D-619)
DV51D-696
RecommendedforusewithIntelliPAP
FlexSet
®
9354D DeVilbiss Serenity
®
9352D
FlexSet Shallow 9354S Serenity Shallow 9352S
FlexSet Gel 9354G Serenity Gel 9352G
FlexSet Gel Shallow 9354GS Serenity Gel Shallow 9352GS
EasyFit
®
Nasal Available in Silicone (S, M, L, XL) or Gel (S, M, L). Contact your equipment provider
EasyFit Full Face Available in Silicone (S, M, L, XL) or Gel (S, M, L). Contact your equipment provider
Electricshockhazard-Donotattempttoopenorremovethecabinet;therearenouser-serviceablecomponents
inside.Ifserviceisrequired,contactyourequipmentproviderforinstructionsonobtainingservice.Openingor
attemptingtoserviceyourdevicewillvoidthewarranty.
Disconnectsupplybeforeservicing.
PressureAccuracy–The IntelliPAP is designed to provide accurate pressure performance from the factory without
additional calibration. No routine service is required provided the device is used in accordance with the manufacturer’s
directions.
NOTE–Some countries require periodic servicing and calibration of this kind of medical device. Contact your equipment
provider for further information.
StandardAir-InletFilter–Check the lter every 10 days and clean it if necessary. Replace the lter every 6 months or
sooner if it is damaged. The standard lter is designed to lter particles larger than 3.0 microns in size. The IntelliPAP
MUST have the standard lter in place during operation.
OptionalFineParticleFilter–Check the lter every 10 days and replace it if it is dirty or damaged; otherwise replace the
lter every 30 days. The optional ne particle lter is designed to lter particles as small as 0.3 microns in size.