Waveform Volume[mL] BreathRate[min-1]
SetPressure
[cmH
2
O]
PressureAccuracy
[Pk-PkcmH
2
O]
Sinusoidal
cycle with I:E =
1:1
500 10, 15, 20
3.0 0.5
7.0 0.5
12.0 1.0
16.0 1.0
20.0 1.0

Test Pressures
3.0cmH
2
O 7.0cmH
2
O 12.0cmH
2
O 16.0cmH
2
O 20.0cmH
2
O
Measured pressure at the patient
connection port (cmH
2
O)
2.1 6.0 11.0 15.0 19.0
Average ow at the patient connection port
(l/min)
70.9 81.5 86.3 90.1 93.1


Breathsperminute 10 15 20
@6.5cmH
2
O
0.5 0.5 0.5
@10cmH
2
O
1.0 1.0 1.0
@13cmH
2
O
1.0 1.0 1.0
@20cmH
2
O
1.0 1.0 1.0
Long Term Static Pressure Accuracy ................................................................................................................. +/- 0.5 cmH
2
O

Warranty .............................................................................................................................. Please contact homecare provider
Equipment classification with respect to protection from electric shock ........................................................................ Class II
Degree of protection from electric shock ................................................................................................. Type BF Applied Part
Degree of protection against ingress of liquids ......................................................................................IPXI Drip-proof vertical
Mode of operation .................................................................................................................................................... Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide


MedicalElectricalEquipmentneedsspecialprecautionsregardingEMCandneedstobeinstalledandputinto
serviceaccordingtotheElectromagneticCompatibility[EMC]informationprovidedintheaccompanying
documents.
PortableandMobileRFCommunicationsEquipmentcanaffectMedicalElectricalEquipment.
Theequipmentorsystemshouldnotbeusedadjacenttoorstackedwithotherequipmentandthatifadjacentor
stackeduseisnecessary,theequipmentorsystemshouldbeobservedtoverifynormaloperationinthe
congurationinwhichitwillbeused.
NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in determining
the suitability of the Equipment or System for the Electromagnetic Environment of use, and in managing the
Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing
other Equipment and Systems or non-medical electrical equipment.