OPERATOR’S MANUAL
mini GAC®
11001-1437-200701
ERROR CODES / 29
Corrective Action
1. Ensure the Loader cup is removed during the empty cell
test.
2. Make sure cell is empty prior to empty cell test.
Pressing the Enter button returns operation to the Home menu.
ERROR 56
Probable Cause
Error 56 displays when the measured average tilt angle exceeds
10 degrees after an Empty Cell and/or a Full Cell test phase.
Corrective Action
1. Hold mini GAC level while performing test or
2. Place on level surface to perform test.
Pressing the Enter button returns operation to the Home menu.