Cu Application
The VenaFlow Elite System will automatically detect the cu that is attached to the
device and apply appropriate pressures. You must attach the cus prior to powering
on the device.
1) Calf cu – Apply cu with the tube pointed toward the foot. The aircell may be
placed either on the back, side or front of the leg. It doesn’t matter where the
aircell sits on the leg. It will perform the same.
Foot cu – Apply cu with the aircell centered on bottom of foot and with the tube
pointed to the left. Foot cus are only for use in standard rapid ination mode.
Thigh cu – Apply cu with the distal and proximal aircells centered on either the
back, side or front of the calf and thigh with the tube pointed toward the foot. It
doesn’t matter where the aircell sits on the leg. It will perform the same.
2) Connect cus to the tubing that is attached to the pump.
For varied patient sizes, trim calf and thigh cus as needed for proper t.
Be sure to secure cu straps snug but not tight. When using the calf cu and thigh cu,
rotation of the aircell on the leg does not aect performance of the system.