EasyManua.ls Logo

DJO VENAFLOW Elite - Specifications; Classifications; Storage;Transportation Conditions; Operating Conditions

DJO VENAFLOW Elite
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
ENGLISH
Specications
Size: 19.3 cm (7.6 in.) x 21.6 cm (8.5 in.) x 11.7 cm (4.6 in.)
Weight: no battery: 1.8 kg (4 lbs.) with battery: 2.2 kg (4.75 lbs.)
Power input: 100 - 240~, 50/60 Hz, 75 VA
Cord: Hospital grade, 4.572 m (15 ft.)
Standards: IEC 60601-1:2005, IEC 60601-1-2:2007
Ination mode: rapid ination and graduated, sequential compression
Bed hanger: accommodates up to 9 cm (3.5 in.)
Battery: LiFePO4 12.8V 1500 mAh
Fuse: F2AL250V, Quantity 2
Classications
Degree of Protection from Electrical Shock: Class 1
Type BF Applied parts (Cus)
IPX0 (Ordinary protection)
No sterilization is required
This device is not intended for use in an Oxygen Rich Environment
Mode of operation is Continuous Operation
Storage/Transportation Conditions
Temperatures: -13° F (-25° C) - 158° F (70° C)
Relative Humidity: 15% - 93%
Atmospheric Pressure: 500 hPa - 1060 hPa
Operating Conditions
Operating Temperatures: 10° C to 40° C
Relative Humidity: 30% to 75%
Atmospheric Pressure: 700 hPa to 1060 hPa
Altitude: below 3000 m

Table of Contents