Cleaning Instructions
The tube assembly is reusable. Both the pump case and the tube assembly can be
cleaned with mild soap solution, antiseptic or disinfectant wipes. Apply using a
dampened cloth or wipes. Do not submerge the device in any liquid.
No adverse eects were found using isopropyl alcohol and hydrogen peroxide (Oxivir
Tb). It is advised to avoid using Dimethyl benzyl ammonium chloride or Dimethyl
ethylbenzyl ammonium cloride (Tec-Surf II), or solutions containing chlorine.
Maintenance
Prior to each use, inspect the device, power cord, cu and tubing for damage.
If damage is detected, contact Customer Service. See Customer Care Contact
Information section.
NOTE: The unit was tested during the manufacturing process and is ready to be placed
into service upon delivery.
Factory Service
When the VenaFlow Elite System requires factory service, contact Customer Service.
See Customer Care Contact Information section.