High Speed (Refrigerated) Micro-Centrifuge
18
2The centrifuge gets into preparation mode and displays last running parameters.
3Press
knob and hold, speed up to reach the maximum speed. While release knob during
acceleration, the centrifuge will start to decelerate and stop.
9. Maintenance
9.1 Cleaning
CAUTIONDo not follow the recommended instructions for cleaning or disinfecting
which may damage the centrifuge.
1Centrifuge
If the centrifuge is exposed to ultraviolet rays for a long time, the color of the doors may be
changed or the label may be came off. After using, cover the centrifuge with a piece of cloth
to protect it from direct exposure.
Once the centrifuge is heavily dirty, clean it with a cloth or sponge moistened with a neutral
detergent solution.
Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
2Rotor chamber
Once the rotor chamber is dirty, clean with cloth or sponge moistened with a neutral detergent
solution. Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
3Drive shaft
Recommend have a regular maintenance for drive shaft. You can wipe the drive shaft with soft
cloth, and then apply a thin coat of silicon grease to it.
4Door
Clean and sterilize the door using the same method as the step 1).
5Rotor
To prevent corrosion, remove the rotor from rotor chamber. If do not use for a lone term, then
CAUTIONDo not directly pour water, neutral detergent or disinfectant solution into
the rotor chamber. Otherwise fluids may leak into the drive units and cause corrosion or
deterioration to the bearings.