3. Required Operational Conditions
3.1 Basic operational conditions
1Power:
D3024: Single-phase, 220V-240V, 50Hz/60Hz, 5A; 110V-120V, 50Hz/60Hz, 5A
D3024R: Single-phase, 220V-240V, 50Hz/60Hz, 10A; 110V-120V, 50Hz/60Hz, 10A
2Ambient temperature: 2-40.
3Relative humidity: 80%.
4No vibration and airflow around.
5No electric dust, explosive and corrosive gases around.
3.2 Transport and storage conditions
1Storage temperature: -40-55.
2Relative humidity: 93%.
4. Installation
This section describes the instructions that you should abide when install the centrifuge to ensure
your safety and the optimum performance. Before moving the centrifuge, the rotor must be removed.
WAR N I NG
Incorrect power supply may damage centrifuge
Make sure the power source conforms to the required power supply before
connecting.
4.1 Location
1Locate this centrifuge on a firmflat and level table, ensure the four feet of this centrifuge stand on
the table firmly. Avoid installing on the slippery table-board that conveys vibration.
2Ideal ambient temperature is 20±5, avoid placing the centrifuge in direct sunlight if temperature
exceeds 30 .
3Ensure the centrifuge has at least 10cm of free space on both sides and at least 30cm behind it to
guarantee cooling efficiency.
4No heat or water resource nearby, otherwise easily cause sample temperature increase or centrifuge