High Speed (Refrigerated) Micro-Centrifuge
22
11. Instructions of rotor and tube
CAUTION
Read the instructions thoroughly, correct use rotor.
Do not exceed the allowable maximum speed of rotortube and adapters etc., be care
that the allowable maximum speed of some adapters are lower than the rotors.
11.1 The rotor instructions
11.1.1 Rotor structure
Figure 11-1 The rotor structure
11.1.2 Available rotors and adapters
All rotors are used for bio-safe when the rotor lid was tightened with the rotor, centrifuge tubes will be
enclosed into rotor to ensure the sample does not leak in centrifugal process. If rotor lid is not available,
the rotor would be no bio-sealing function. The rotors can be used as follows:
Table 11.1 Rotors and adapters
Rotor
type
ID
code
Tube/bottle
Adapter
Maximun
speed
(rpm)
Maximum
centrifugal
radius
r
max
(cm)
Maximum RCF
Rcf (×g)
1 AS24-2 2/1.5ml tube 15000 8.5 21380
Knob Locking nut Cover
Seal ring Shaft hole