High Speed (Refrigerated) Micro-Centrifuge
19
detach the rotor door and upside the rotor down to dry the tube holes.
If a sample is leaked in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to
the rotor when it is completely dried.
The rotor should be regular maintenance, recommend to cleaning it each 3 months to ensure tube
and rotor holes keep clean, and then apply a thin coat of silicon grease.
6Drain (D3024R)
The centrifuge is equipped with drainage, drain off water when more water is in drain pipe.
9.2 Consumables
The table list below is the centrifuge consumables. It is recommended to replace consumables
according the table below.
No. Consumables Replacement conditions
1 Seal ring of door
Crack
2 Rubber block of temperature sensor
3 Seal ring of centrifuge chamber (3024R)
9.3 The replacement of seal rings
9.3.1 Instructions
There are three high-temperature rubber seal rings that equipped into rotor to achieve bio-safe. The seal
rings may fall off or aging after several autoclaving, need to be replaced or re-installed.
Figure 9-1 Seal rings of rotor
9.3.2 Replacement methods
1Clean the seal ring slot with neutral detergent solution and make it dry.
2Evenly coated with glue (501) in the seal ring slot and keep the seal ring into slot, press evenly to
Seal ring 1 2 3
Seal ring 2 3