High Speed (Refrigerated) Micro-Centrifuge
24
Cleaning fluids
Alkalinehigher than pH9 O X O
AlkalinepH9 or lower O O O
NeutralpH7 O O O
Warm water(up to 70 ) O O O
Ultrasonic
cleaning
Neutral detergentpH7 O O O
Sterilization
Autoclaving
1150.7kg/cm
2
30minutes O O O
1211.0kg/cm
2
20 minutes X O O
1261.4kg/cm
2
15 minutes X X X
Boiling 15 to 30 minutes O O O
Ultraviolet
sterilization
200-300nm X X X
Gas sterilization Ethylene oxide O X O
Formaldehyde O O O
PA: Polyallomer PC: Polycarbonate PP: Polypropylene
11.2.2 Cleaning PC tubes
PC materials are low in chemical resistance against alkaline solutions. Avoid using neutral detergents
with pH higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted
according to the instruction in the makers catalog. Use detergent with its pH between 7 and 9.
11.2.3 Autoclaving PAPC and PP tubes
PA begins softening at about 120 , PC and PP at about 130℃℃. Autoclave PA tubes at 1150.7kg/cm
2
for 30 minutes and PC and PP tubes at 1210.1kg/cm
2
for 20 minutes. If a certain temperature is
exceeded, the tubes may be deformed.
When use a sterilizing chamber, please operate as follows:
1Place tubes in vertical position, mouths upward. If tubes are placed sideways, they may deform
into an oval shape due to gravity.
2Remove screw nuts and inner doors to prevent from deformation or rupture.
3Wait until the sterilizing chamber cools down to the room temperature before the tubes are
removed.
11.2.4 Condition and life expectancy of tubes
The life expectancy of plastic tubes depends on the characteristics of samples, speed of the rotor used, and
temperature applied, and so on. When the plastic tubes are used for centrifuge of ordinary aqueous samples
(pH between 5.0 and 9.0), their life expectancies are defined as follows.
Be operated at the maximum speed:
High quality tubes (PAPCPP): 30-50 operations