1. Safety instructions
Pay attention to the instructions for use when handling the lamp.
WARNING:
This device has not been designed for use in potentially explosive areas.
According to the Medical Device Regulation the light is classified under class I.
Store the OT-light in its package for at least 24 hours in the respective room before
mounting, in order to equal temperature differences.
Please read the instructions for use carefully to make the most of your lighting
system and to avoid any damages to the device.
The lights may only be repaired and special assembly work may only be carried
out on the reflector or sockets by ourselves or a company that has been expressly
authorized by us.
The manufacturer can only be made responsible for the safety of the light if repairs
and alterations are carried out by the manufacturer himself or a company that
guarantees to observe the safety regulations.
The manufacturer cannot be made liable for personal or material damages if the
light is operated inexpediently or incorrectly or used for purposes other than those
for which it is intended.
The light is to be dismantled from the spring arm in reverse order to its assembly.
This may only be carried out after the spring arm has been adjusted in height at
horizontal position since the arm is under spring tension and can bounce up.
Make sure that the light is in perfect working order before every use.
Attention, external power supply!
The light works only with an external power supply 250VA.
The external power supply used with the OT-light must be tested and validated
according to IEC 60601-1.
Attention!
A main control switch must be installed for turning the system power-off.
ATTENTION!
During the mounting of the OT-lights the entire system
(incl. the ceiling attachment) must be disconnected from mains!
A later dismounting of the lights from the spring arms or dismount-
ing the sliding contacts inside the arms is to be done ONLY AFTER
DISCONNECTING THE ENTIRE SYSTEM FROM MAINS.
Otherwise the main control board will be damaged!