Performing the measurement
Automatic Measurement
The sampling is automatically triggered after attaining the minimum breath volume and the
minimum blowing period.
Preparation
Procedure
Switch on the device (see page 44).
Insert a new mouthpiece in the mouthpiece receptacle
(see page 45).
Af
ter 2 seconds the following appear in the display:
customer specific info (if activated),
or
Input of data, see "Data input before or after the measurement" on
page 47
or
» WAIT «.
The customer specific info can also be accessed by pressing the
key. End with the
key.
After a temperature-dependent heating period, the following appears
in the display: » READY « and a short audible signal beeps. The cur-
rent test number is displayed at the same time. The right LED flashes
green.
The device is ready for measurement.
Prompt the test person to blow evenly and continuously into the
mouthpiece. A sufficient breath flow is confirmed by a continuous
audible signal and the green LED.
Dur
ing the delivery of the breath sample the following message
appears in the display: » BLOW «.
If the breath sample is sufficient (minimum breath volume and the min-
imum blowing period were reached), the sampling is triggered. Then
the green LED goes out, the audible signal stops.
The following message appears in the display:
» WAIT Analyzing «
01023996_01_en.eps
11:53:47
04.12.2007
Test No. 26
READY
OK
01123996_01_en.eps
11:53:47
04.12.2007
BLOW