Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 139
Tests
8.4.2 Gas flow
8.4.2.1 Gas cylinders
1. Open the gas cylinder valves slowly.
Check that the displayed pressures are sufficient.
When using pressure reducers without electronic pressure measurement, read
the pressure from the pressure gauge.
2. Close the gas cylinder valves.
On devices that are equipped with Advanced Cylinder Support, the gas cylinder
valves can remain open during operation. These devices are identified by an
appropriate label near the gas inlets (see "Gas inlets", page 26).
8.4.2.2 Checking the emergency O
2 delivery (with electronically controlled gas
mixer)
Prerequisite: Y-piece is sealed.
1. Set the O
2 switch (2) to the Add. O2 position.
3510935110
1
2