Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 283
Technical data
Information Gas consumption related to O2 flush and O2
therapy is not included in this calculation.
Accuracy ±15 % or ±2 L/min (the larger value applies)
Resolution of displayed value 1 L
Anesthetic agent consumption
Range 0 to 999.9 mL fluid
Accuracy ±25 % or ±1 mL (the larger value applies)
Resolution of displayed value 0.1 mL
Waveforms O
2 concentration
Primary anesthetic gas concentration
CO
2 concentration
Airway pressure
Volume (only for loops)
Flow
Sweep speed 6.25; 12.5; 25.0 mm/s
Scale
Airway pressure –20 to 80 hPa (cmH
2O)
Flow –120 to 120 L/min
Volume 0 to 2000 mL
O
2 0 to 100 Vol%
CO
2 0 to 100 mmHg (0 to 12 Vol%, 0 to 12 kPa)
Anesthetic agent
Halothane 0 to 5 Vol% (kPa)
Enflurane 0 to 6 Vol% (kPa)
Isoflurane 0 to 5 Vol% (kPa)
Sevoflurane 0 to 10 Vol% (kPa)
Desflurane 0 to 20 Vol% (kPa)
Loops Pressure-Volume
Flow-Volume
16.18 Operating characteristics
Mains power supply
Mains voltage 100 to 240 V AC at 50/60 Hz
Maximum power consumption 4 A
Power cable
Maximum length 5 m (16.4 ft)
Protective ground resistance Maximum 0.1 Ω
16.17 Display of calculated values (continued)