284 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Operating voltage 250 V
Operating current 10 A
Current consumption at 230 V AC
Standby (without charging the internal battery) <0.18 A
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2, PEEP =
0 hPa, FG flow 4 L/min O
2) without charging the
internal battery
<0.40 A
Maximum 2 A
Current consumption at 110 V AC
Standby (without charging the internal battery) <0.38 A
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2, PEEP =
0 hPa, FG flow 4 L/min O
2) without charging the
internal battery
<0.84 A
Maximum 4 A
Power consumption
Standby (without charging the internal battery) <40 W
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2, PEEP =
0 hPa, FG flow 4 L/min O
2) without charging the
internal battery
<95 W
Maximum 400 W
Peak inrush current Approx. 8 to 14 A
Approx. 6 to 10 A quasi-RMS
Internal battery
Type Lead-gel battery
Sealed, maintenance-free
Capacity 7.2 Ah
Voltage 24 V
Fuse F15A 80V UL248-14, breaking capacity
1000 A, size 19.7 mm x 19 mm x 5 mm
Current Maximum 15 A
Backup time with new and fully charged battery
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2,
PEEP = 0 hPa, FG flow 4 L/min O
2)
At least 45 min
Typically 120 min
During ventilation in Man / Spon At least 90 min
Charging time (to achieve at least 30 minutes’
backup time)
At least 8 hours
16.18 Operating characteristics (continued)