Technical data
Charging power Maximum 50 W
Gas supply
Gas quality
Oil content <0.1 mg/m
3
Dew point 5 °C (41 °F) at ambient temperature
Particle size Dust-free air (filtered with pore size <1 μm)
Supply pressure for O
2, Air, N2O 2.7 to 6.9 kPa x 100
39 to 100 psi
0.27 to 0.69 MPa
Maximum short-term peak inlet flow at 6.9 kPa
x 100 (100 psi or 0.69 MPa) supply pressure
O
2 135 L/min (applies only when there is no distri-
bution piece for the central O
2 supply)
Air
Without bronchial suction system 50 L/min
Including a directly connected bronchial
suction system
130 L/min
N
2O40L/min
Drive gas Not needed
Gas supply connection Depending on configuration: NIST, DISS
(CGAV-5/B or CGAV-5/N), French standard
(NFS90-116)
Gas cylinders (dimensions)
Diameter 100 to 140 mm (3.94 to 5.51 in) for versions
with upright gas cylinders
100 to 102 mm (3.94 to 4.01 in) for versions
with hanger yoke system for gas cylinders with
pin-index connections
Maximum height 880 mm (34.64 in) for versions with upright gas
cylinders
757 mm (29.80 in) for versions with hanger
yoke system for gas cylinders with pin-index
connection
Pressure reducers
Version Conforms to DIN EN ISO 10524-1
Permissible inlet pressure (P
V)
Air, O
2 Up to 200 kPa x 100 (2900 psi, 20 MPa)
N
2O Up to 60 kPa x 100 (870 psi, 6 MPa)
16.18 Operating characteristics (continued)