Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 149
Tests
Seal the sensor port with the sealing cap.
5. Follow the instructions on the screen.
The device will perform the calibration.
6. After the calibration is complete, remove the sealing cap and plug it back in the
holder.
Insert the O
2 sensor (4) back into the sensor port and turn it clockwise until it
reaches the end position.
7. After the calibration, the fully automatic main test is continued.
Note on the expiry of the life span of the O
2 sensor cell
The life span of the O
2 sensor cell is 2 years and is monitored by the device. As
soon as there are 28 days or fewer of the life span left, the remaining life span will
be displayed in the system test. When the life span has expired, a corresponding
message and the following illustration will be displayed: