206 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
!! 30 Battery charge very low The battery charge is critical
and the mains power supply
is not available. The device
will shut down in the next 5
minutes.
Make sure that the mains
power supply is correctly
connected.
The breathing system
warmer has been switched
off. Check the breathing cir-
cuit for condensate. Increase
the fresh-gas flow if neces-
sary.
Once the battery has been
depleted, ventilate the patient
manually.
!! 170 Battery failure The battery is faulty. If the
mains power supply fails, the
device will switch off immedi-
ately.
Call Dräger.
! 100 Battery temperature high The battery temperature is
high. Charging of the battery
has been suspended to pro-
tect it from damage.
Ensure that the system is
connected to the mains
power supply.
Check the ambient tempera-
ture.
! 100 Breathing circuit expired Accessory has been used
too long.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
!!! 120 Breathing system failure The breathing system is not
correctly assembled or
installed.
Check patient condition.
Check the breathing system.
Verify that the sensor port for
the O
2 sensor is sealed.
Check if the piston dia-
phragm is correctly inserted.
If the problem persists,
replace the breathing sys-
tem.
The breathing system is
faulty.
Replace the breathing sys-
tem.
If ventilation is impaired, use
a manual resuscitator.
Priority Alarm Cause Remedy