EasyManua.ls Logo

Dräger Atlan A300 - Page 209

Dräger Atlan A300
332 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 209
Troubleshooting
!! 150 Emergency air inlet acti-
vated
There was not enough gas to
ventilate the patient. To main-
tain a minimum ventilation,
the device has started to use
ambient air.
Refill the breathing bag, e.g.,
with O
2 flush.
Increase fresh-gas flow.
Check the breathing circuit
for tight connections and
leakages.
An external anesthetic gas
monitor is extracting gas
from the breathing system
via a sample line.
Disconnect the sample line.
Seal the sample line port.
!! 135 etCO
2 high etCO2 has exceeded the
upper alarm limit.
Check ventilation.
!! 135 etCO
2 low etCO2 is below the lower
alarm limit.
Check ventilation.
!! 100 Exp. pressure sensor fail-
ure
Sensor calibration failed. Ensure that a suitable substi-
tute monitoring is available.
Perform the system test.
If the problem persists, call
Dräger.
! 100 Expiratory flow sensor
expired
Accessory has been used
too long.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
acknowledge alarm.
! 255 Expiratory flow sensor
failure
The sensor is not calibrated.
The breathing system has
been replaced or discon-
nected since last calibration.
Perform the leakage test.
Failure of the flow sensor. Replace the flow sensor and
perform a leakage test.
!!! 200 External fresh-gas outlet
failure?
The internal fresh-gas valve
is faulty. It is unclear whether
the fresh gas is being fed into
the internal breathing sys-
tem or towards the external
fresh-gas outlet.
Use "O
2+" button to deter-
mine flow direction of fresh
gas: -If internal breathing
system or breathing bag fill,
external fresh-gas outlet is
not available. -If gas flows
out of the external fresh-gas
outlet, external fresh-gas out-
let can be used. Internal
breathing system can only be
used when the breathing bag
is not connected (ventilation
with ambient air only).
Check fresh-gas settings.
If the problem persists, call
Dräger.
Priority Alarm Cause Remedy

Table of Contents