234 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
4. Turn the knurled nut (4) approximately 90° counterclockwise.
5. Remove the sensor cap (5).
Do not remove the sensor cell. Otherwise, the calibration data will be lost.
This may result in a waiting time when putting the device back into operation.
6. WARNING
Risk of incorrect gas measurement
Reprocessing and sterilization can damage the O
2 sensor cell. This may result
in a malfunctioning measurement and the patient may be put at risk.
Do not disinfect or sterilize the O2 sensor cell, or immerse it in other fluids.
Place the O
2 sensor as shown (6).
The sealing cap (7) remains on the holder (8).