Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 239
Reprocessing
13.9.2 Disassembling the components
13.9.2.1 CLIC adapter (option)
1. Screw off the CLIC adapter (1).
2. Press the release button (2). The holder (3) flips open.
3. Push the cleaning plate (4) fully into the adapter.
4. Close the holder (3) so that it engages.
13.9.2.2 Flow sensors and ports
CAUTION
Risk of failure of flow measurement
Improper reprocessing and soiling, such as residues or particles, can damage the
flow sensor.
No machine cleaning or disinfection
No plasma sterilization or radiation sterilization
No water jets, compressed air, brushes or the like
No ultrasonic bath
No hot-steam sterilization with Spirolog and Infinity ID flow sensors
Clean and disinfect the flow sensor in accordance with the corresponding
instructions for use.
For disinfecting the flow sensor use only clean disinfectant solutions.