278 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Accuracy without PGM ±50 % of the measured value or ±5 hPa/L/s
(cmH
2O/L/s) (the larger value applies)
Information The values may be distorted by spontaneous
breathing, thus reducing the measurement
accuracy.
Resolution of displayed value 1 hPa/L/s (cmH
2O/L/s)
Elastance E
Range 0.005 to 10 hPa/mL (cmH
2O/mL)
Resolution of displayed value 0.001 hPa/mL (cmH
2O/mL)
Supply pressures
Central supply
Range 0 to 9.8 kPa x 100
0 to 140 psi
0 to 0.98 MPa
Accuracy (up to 7 kPa x 100) ±4 % or ±0.2 kPa x 100 (the larger value
applies)
±4 % or ±3 psi (the larger value applies)
±4 % or ±0.02 MPa (the larger value applies)
Resolution of displayed value 0.1 kPa x 100
1psi
0.01 MPa
Gas cylinders
Range 0 to 250 kPa x 100
0 to 3600 psi
0 to 25 MPa
Accuracy ±4 % or ±6 kPa x 100 (the larger value applies)
±4 % or ±87 psi (the larger value applies)
±4 % or ±0.6 MPa (the larger value applies)
Resolution of displayed value 1 kPa x 100
1psi
0.1 MPa
Results in the system test
Total leakage
Range 10 to 5000 mL/min (measured at BTPS, refer-
enced to a pressure of 30 hPa)
Accuracy ±25 % or ±150 mL/min (the larger value
applies)
Resolution of displayed value 1 mL/min
Leakage during mechanical ventilation
Range 10 to 1000 mL/min (measured at BTPS,
referenced to a pressure of 30 hPa)
16.13 Measuring systems and displays (continued)