Instructions for use Fabius GS premium SW 3.n 169
Cleaning, disinfection and sterilization
Removing the compact breathing system
Before removing the compact breathing system,
the following hoses and cables must be removed:
Flow sensor cable
–O
2 sensor cable and O2 sensor capsule
Pressure measurement hose
APL bypass hose
PEEP/PMAX hose
Removing the inspiratory valve
1 Remove the valve cover plug (B) or the O2
sensor from the dome (C) of the inspiratory
valve.
2 Screw off the cap nut (A).
3 Remove the dome (C).
4 Take out the valve plate (D).
5 Remove the sealing ring (E) from the socket
(F).
Removing the expiratory valve
1 Screw off the cap nut (G).
2 Remove the sight glass (H).
3 Take out the valve plate (I).
4 Remove the sealing ring (J) from the socket
(K).
Removing the exhaust port
Screw off the exhaust port (A).
20980
A
B
C
D
E
F
G
H
I
J
K
20994