Instructions for use Fabius GS premium SW 3.n 181
Maintenance
Inspection
Inspections must be carried out regularly
according to the following guidelines and within the
specified intervals. Technical documentation is
available on request.
Safety checks
The safety checks are no substitute for service
measures indicated by the manufacturer, including
the preventive replacement of wearing parts.
1 Check the accompanying documents:
Latest instructions for use are available
2 Check the following functions according to the
instructions for use:
Check the proper function of the flow
measurement.
Check the function of the pressure
measurement based on the parameters
PAW, PEEP, PMAX.
Check the proper function of the O
2
measurement.
Check the function of the anesthetic
vaporizer according to the associated
instructions for use.
Check the function of the O
2 flush.
Check the function of the pressure reducer
(optional) of the compressed gas cylinder.
3 Check if the product is in good condition:
Labels are complete and legible
No visible damage
Fuses which are accessible from the
outside are in compliance with the specified
values
Country-specific labeling of gas types
4 Check components and accessories for
completeness according to the instructions for
use.
5 Check for electrical safety in compliance with
IEC 62353.
6 Check safety features:
Check the functional state of the optical and
acoustic alarm generators.
Check the functional state of the O
2 failure
alarm.
Check the locking device of the anesthetic
vaporizer.
Check the function of the power failure
alarm and the battery function.
Check S-ORC functionality.
Checks Interval Responsible personnel
Inspection and safety checks
1)
1) Designation applies to the Federal Republic of Germany; corresponds to the “Recurring safety inspection” in the Federal
Republic of Austria
Every 12 months Experts
WARNING
Risk of malfunction of the medical device
If the safety checks are not regularly
performed, the medical device might not
function properly.
Perform the safety checks in the specified
intervals.