Dräger FPS-COM 7000 17
Maintenance
5 Maintenance
5.1 Maintenance intervals
Insert cannot be fitted/
pulled out
Screw blocks the
insert
Lift screw.
Insert path blocked Check sealing and
grease or replace if
necessary.
Clear insert path and
carefully fit insert.
Screw not completely
loosened
Completely loosen
the screw, making
sure that the spring
below the screw can
move freely, lift it if
necessary.
Communication unit
cannot be fitted/
dismantled
Fixing lugs on
facepiece visor frame
not available or
defective
Replace visor frame.
Mount defective Replace mount.
Fastening eyes on
communication unit
defective
Replace fastening
eyes.
Communication unit
not hooked in
correctly
Dismantle
communication unit
and fit again.
Earphone cannot be
bent to desired
position
Earphone deformed
or gooseneck
defective
Replace earphone.
Fault Cause Remedy
No connection via
cable to radio
possible
Cable defevtive Replace cable.
Cable not fitted
correctly
Check connection
between cable and
radio. Fit cable again
if necessary.
Check connection
between cable and
communication unit.
Fit cable again if
necessary.
Incorrect cable
connected
Use suitable cable.
Push-button jams Push-button soiled Clean push-button.
Slide switch jams Slide switch soiled Dismantle and clean
slide switch.
Slide switch damaged Replace slide switch.
Work to be carried out
Yearly
Check the functioning of the communication unit X
Fault Cause Remedy