DRÄGER INFANT INCUBATOR C2000/C2000E
VF8 DELTA/DELTA XL/KAPPA 28-25
Quick Reference Table – C2000/C2000e Setup
To access or execute C2000/C2000e functions listed below, select a menu item, scroll
to the appropriate setting and click
The Incub1 presents the following selections:
Menu Item Description Available Settings
Show All
Parameters
Displays all parameters
# Parameter
Boxes
Select number of Parameter Boxes 1 (default), 2
NOTE:
• Incub2 Parameter box is not displayed
on screen or in parameter priority list
when the number of Parameter Boxes
is set to 1
• When set to 2, the parameter box must
be selected from the pick list to be
displayed in the appropriate channel
Parameter 1 Selects up to 3 parameters for
display in Incubator parameter box
T1skin (default), T2skin, Tair, %O2,
R.H., Wt, None
Parameter 2 T1skin, T2skin, Tair, %O2, R.H.
(default), Wt, None
Parameter 3 T1skin, T2skin, Tair (default), %O2,
R.H., Wt, None
Incubator
Mini Trend
Allows you to select parameters for
display in mini trends (See page 6-
7 for more information on mini
trend display).
NOTE: Both Incub1 & Incub2 can
display a mini trend.
T1skin (default), T2skin, Tair, %O2,
R.H., Wt
Wt This is a read-only field. Only the
time, month, and day appears with
the Weight label.
Not applicable